RECHARGEABLE LEAD ACID BATTERIES PRISMAFLEX FG20121; 12V 1.2AH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-04-09 for RECHARGEABLE LEAD ACID BATTERIES PRISMAFLEX FG20121; 12V 1.2AH manufactured by Gambro; Fiamm.

Event Text Entries

[1949910] Prismaflex premature battery failure-potential for harm: four power backup batteries, have failed the biomed battery analyzer test, belong to four seperate devices. These batteries were part of the original purchased devices, we acquired less than 2 years ago. Additionally the manufacturer does not state clearly, in their maintenance instructions a required frequency to inspect the battery.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2065684
MDR Report Key2065684
Date Received2011-04-09
Date of Report2011-04-09
Date of Event2011-03-07
Report Date2011-04-09
Date Reported to FDA2011-04-09
Date Added to Maude2011-04-21
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameRECHARGEABLE LEAD ACID BATTERIES
Generic NameCRRT
Product CodeFCO
Date Received2011-04-09
Returned To Mfg2011-03-25
Model NumberPRISMAFLEX
Catalog NumberFG20121; 12V 1.2AH
Lot Number03089
ID Number*
Device AvailabilityR
Device Age2 YR
Device Sequence No1
Device Event Key0
ManufacturerGAMBRO; FIAMM
Manufacturer Address14143 DENVER WEST PARKWAY LAKEWOOD CO 80401 US 80401

Device Sequence Number: 2

Brand NameRECHARGEABLE LEAD ACID BATTERIES
Generic NameCRRT
Product CodeFCO
Date Received2011-04-09
Returned To Mfg2011-03-25
Model NumberPRISMAFLEX
Catalog NumberFG20121; 12V 1.2AH
Lot Number04089
ID Number*
Device AvailabilityR
Device Age2 YR
Device Sequence No2
Device Event Key0
ManufacturerGAMBRO; FIAMM
Manufacturer Address14143 DENVER WEST PARKWAY LAKEWOOD CO 80401 US 80401

Device Sequence Number: 3

Brand NameRECHARGEABLE LEAD ACID BATTERIES
Generic NameCRRT
Product CodeFCO
Date Received2011-04-09
Returned To Mfg2011-03-25
Model NumberPRISMAFLEX
Catalog NumberFG20121; 12V 1.2AH
Lot Number04089
ID Number*
Device AvailabilityR
Device Age2 YR
Device Sequence No3
Device Event Key0
ManufacturerGAMBRO; FIAMM
Manufacturer Address14143 DENVER WEST PARKWAY LAKEWOOD CO 80401 US 80401


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-09

© 2026 FDA.report