OLYMPUS VIDEO TELESCOPE WA50011A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-04-15 for OLYMPUS VIDEO TELESCOPE WA50011A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[21676497] The user facility reported that during a laparoscopic cholecystectomy, the users experienced a complete loss of image. The procedure was completed using a different device. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[21869584] The device referenced in this report was returned to olympus for eval. The eval confirmed the user's report of image lost. There were multiple dents and a slight bend on the outer tube. There were scratches, dent on the frame. The video connector contact pins were corroded/burnt. The video connector was cracked. These phenomena were attributed to physical damage. The device was serviced and was returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2011-00013
MDR Report Key2068787
Report Source06
Date Received2011-04-15
Date of Report2011-03-17
Date of Event2011-03-07
Date Mfgr Received2011-03-17
Date Added to Maude2011-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameTELESCOPE
Product CodeFBP
Date Received2011-04-15
Returned To Mfg2011-03-17
Model NumberWA50011A
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.