MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-04-12 for ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN 806516100 manufactured by Bio-rad Medical Diagnostics Gmbh.
[1945237]
The customer complained false negative reactions of a proficiency sample (aut-02) containing anti-e. We received the proficiency sample (aut-02) but not the complained lot of anti-human globulin, anti-igg solidscreen ii. The proficiency sample was tested with the retention sample on tango in the quality control laboratory and reacted negatively. Furthermore, a twofold serial dilution was tested on solidscreen ii and reacted correctly positive. Referred to informations of customer the proficiency sample aut-02 contains a monoclonal anti-e and not a natural occurring antibody. The intended use of solidscreen is, amongst others, the detection of irregular red cell antibodies in human blood. The correct function of the affected lot anti-human globulin, anti-igg solidscreen ii was confirmed by testing different weak antibodies. All positive and negative reactions were correct.
Patient Sequence No: 1, Text Type: D, B5
[9105699]
Stn #: 125098.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9610824-2011-00043 |
MDR Report Key | 2070595 |
Report Source | 05,06 |
Date Received | 2011-04-12 |
Date of Report | 2011-04-12 |
Date of Event | 2011-02-08 |
Date Mfgr Received | 2011-03-14 |
Date Added to Maude | 2012-02-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | DR MARC GORZELLIK, HEAD OF |
Manufacturer Street | INDUSTRIESTRASSE 1 |
Manufacturer City | DREIEICH 63303 |
Manufacturer Country | GM |
Manufacturer Postal | 63303 |
Manufacturer Phone | 103801528 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN |
Generic Name | ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN |
Product Code | DEW |
Date Received | 2011-04-12 |
Catalog Number | 806516100 |
Lot Number | 7016110-05 |
Device Expiration Date | 2011-10-12 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIO-RAD MEDICAL DIAGNOSTICS GMBH |
Manufacturer Address | INDUSTRIESTRASSE 1 DREIEICH D-63303 GM D-63303 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2011-04-12 |