ACCESS THYROGLOBULIN ANTIBODY II A32898

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 02,07 report with the FDA on 2011-05-02 for ACCESS THYROGLOBULIN ANTIBODY II A32898 manufactured by Beckman Coulter Inc..

Event Text Entries

[1851898] This cf was opened to capture further evaluation of thyroglobulin antibody assay (thgaby) not meeting linearity specifications. In 2009, beckman coulter inc. , (bci) was performing internal dxi onboard dilution linearity studies for multiple assays. Wb ii, sample diluent a, and s0 calibrators were processed during this study to qualify wb ii as an additional assay diluent. Wb ii, sample diluent, and s0 calibrators did not meet the assay performance criteria linearity specifications. No patient results were generated in this event. Patient treatment was not affected.
Patient Sequence No: 1, Text Type: D, B5


[9044385] This event will be further evaluated and investigated. Note: this reportable event was identified during a retrospective review on 04/01/2011 of complaints for additional reportable events. This cf is related to (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2122870-2011-01239
MDR Report Key2074294
Report Source02,07
Date Received2011-05-02
Date of Report2009-12-01
Date of Event2009-12-01
Date Mfgr Received2009-12-01
Date Added to Maude2012-02-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street1000 LAKE HAZELTINE DRIVE
Manufacturer CityCHASKA MN 55318
Manufacturer CountryUS
Manufacturer Postal Code55318
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCESS THYROGLOBULIN ANTIBODY II
Generic NameACCESS THYROGLOBULIN ANTIBODY II
Product CodeJNL
Date Received2011-05-02
Model NumberNA
Catalog NumberA32898
Lot Number831065
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address1000 LAKE HAZELTINE DRIVE CHASKA MN 55318 US 55318


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-02