WATERS TQD WITH MASSLYNX SOFTWARE TDQ 186003875

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-04-15 for WATERS TQD WITH MASSLYNX SOFTWARE TDQ 186003875 manufactured by Waters Corporation.

Event Text Entries

[1947837] This (b)(6) laboratory customer reported results of pt samples (analyte benzodiazepines concentration assay) indicating a higher analyte concentration than the results calculated by waters' mass spectrometry software, targetlynx. The reported results were exported from (b)(6) to the customer's laboratory information management system (lims). Consequently, incorrect results (false positives) for fourteen (14) pts were reported by this laboratory to its customers. The laboratory subsequently identified the discrepancy and reported the correct results to its customers within twenty-four (24) hours. Repeated attempts by waters to gather information related to any pt impact from incorrect results being reported have been unsuccessful.
Patient Sequence No: 1, Text Type: D, B5


[9043982] The investigation into this issue indicated the laboratory had downgraded its (b)(6) software to a version that requires manual verification of data exported to a lims. The laboratory failed to execute its required manual verification of exported data. The customer subsequently performed a manual check of the (b)(6) data file for the affected pt samples and confirmed the results were accurately reported by (b)(6) and that these values differed after xml exportation to its lims. The customer also started that a review of historical data was performed and no other incidents of this error were identified. Should additional information become available, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616673-2011-00002
MDR Report Key2074577
Report Source06
Date Received2011-04-15
Date of Report2011-04-15
Date of Event2010-06-15
Date Mfgr Received2010-06-15
Device Manufacturer Date2008-11-01
Date Added to Maude2011-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHLEEN MORAHAN
Manufacturer Street34 MAPLE STREET
Manufacturer CityMILFORD MA 01757
Manufacturer CountryUS
Manufacturer Postal01757
Manufacturer Phone5084823574
Manufacturer G1WATERS TECHNOLOGIES IRELAND, LTD.
Manufacturer StreetWEXFORD BUSINESS PARK, DRINAGH
Manufacturer CityWEXFORD
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWATERS TQD WITH MASSLYNX SOFTWARE
Generic NameMASS SPECTROMETER FOR CLINICAL USE (DOP)
Product CodeDOP
Date Received2011-04-15
Model NumberTDQ
Catalog Number186003875
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWATERS CORPORATION
Manufacturer AddressMILFORD MA US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.