MICROSMART

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-04-14 for MICROSMART manufactured by Ge Healthcare.

Event Text Entries

[21349923] The month of manufacture is not known at this time. This device was originally manufactured in (b)(4) but the facility no longer manufactures devices.
Patient Sequence No: 1, Text Type: N, H10


[21410859] It was reported that a nurse was performing an electrocardiograph (ecg) when she smelled a faint burning scent. She went on to perform 2 more ecgs and the smell increased. The unit was switched off and a technician was called to check the device. The technician turned to unit back on 5 minutes later and noticed visible smoke coming from the device. The unit was switched off and the battery removed. The battery was reported to be near room temperature. After a few minutes, a blister formed on the plastic housing near the ecg cable connection. The blister developed into a blackened hole. The unit's motherboard was visibly hot. There were no injuries reported. Ge healthcare's investigation is ongoing. A follow up report will be submitted once the investigation has been completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2124823-2011-00041
MDR Report Key2081135
Report Source01,05,06
Date Received2011-04-14
Date of Report2011-04-14
Date of Event2011-03-16
Date Mfgr Received2011-03-16
Device Manufacturer Date1998-01-01
Date Added to Maude2011-07-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOY SONSALLA
Manufacturer Street3000 N. GRANDVIEW BLVD. W450
Manufacturer CityWAUKESHA WI 53188
Manufacturer CountryUS
Manufacturer Postal53188
Manufacturer Phone2625482661
Manufacturer G1GE HEALTHCARE
Manufacturer CityMILWAUKEE WI
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICROSMART
Generic NameECG ANALYSIS SYSTEM
Product CodeLOS
Date Received2011-04-14
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerGE HEALTHCARE
Manufacturer AddressMILWAUKEE WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.