3M RED DOT NEONATAL MONITORING ELECTRODES 2269T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,07 report with the FDA on 2011-04-13 for 3M RED DOT NEONATAL MONITORING ELECTRODES 2269T manufactured by 3m Health Care.

Event Text Entries

[1983579] A 3m (b)(4) stated the customer reported that when the (b)(4) 3m red dot neonatal monitoring electrodes are plugged into the lead, the machine states "left leg off". Replacement product was sent to the sales representative. The affected sample was sent to 3m and analyzed. Testing found that the defect was isolated to the red wires from mn wire lot 100827. Mn wire has been contacted and will investigate internally to determine a root cause and how much of the lead wire inventory is affected. 3m will continue to monitor for this defect. Corrective action has been recommended.
Patient Sequence No: 1, Text Type: D, B5


[9122911] Patient information was not available/no patient(s) involved in the complaint. No adverse patient effects were reported. If additional information is received, a follow-up report will be submitted. Electrical wire.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2110898-2011-00039
MDR Report Key2083181
Report Source04,07
Date Received2011-04-13
Date of Report2011-03-15
Date of Event2011-02-02
Date Mfgr Received2011-03-15
Device Manufacturer Date2010-08-01
Date Added to Maude2011-05-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANDREA GRABANSKI
Manufacturer Street3M CENTER, BLDG. 275-5W-06
Manufacturer CityST. PAUL MN 55144
Manufacturer CountryUS
Manufacturer Postal55144
Manufacturer Phone6517331047
Manufacturer G13M CANADA
Manufacturer Street400 RT 100
Manufacturer CityMORDEN, MANITOBA R6M1Z9
Manufacturer CountryCA
Manufacturer Postal CodeR6M 1Z9
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M RED DOT NEONATAL MONITORING ELECTRODES
Generic Name870.2360 ELECTROCARDIOGRAPH ELECTRODE
Product CodeFPX
Date Received2011-04-13
Model NumberNA
Catalog Number2269T
Lot Number2012-08 PJ
ID NumberNA
Device Expiration Date2012-08-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer AddressST. PAUL MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-13