LIPASE OSR6230

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2011-05-12 for LIPASE OSR6230 manufactured by Beckman Coulter Biomedical Limited.

Event Text Entries

[1979363] A customer contacted beckman coulter inc. (bci) in regards to an erroneously high lipase result of 461. 0 u/l generated by (b)(4) chemistry analyzer. The reaction monitor for the instrument displayed a non-typical curve. The result was not reported out of the laboratory. Subsequent testing on an alternate instrument produced a lower result of 67. 45 u/l and the reaction monitor displayed a typical curve. There was no effect to patient treatment.
Patient Sequence No: 1, Text Type: D, B5


[9107603] There was no issue with control recovery and no other samples were affected at this time. The customer indicated that this issue re-occurs every few weeks. The customer did not retain the sample for further investigation. The customer has been requested to retain any further samples which demonstrate atypical lipase values for investigation. A copy of hard drive data was requested for investigation. The sampling order and contamination parameters have been changed by the laboratory. Since these changes were implemented no more atypical lipase values have been obtained.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-01729
MDR Report Key2087736
Report Source01,06
Date Received2011-05-12
Date of Report2011-03-28
Date of Event2011-03-09
Date Mfgr Received2011-03-28
Device Manufacturer Date2010-06-01
Date Added to Maude2012-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER BIOMEDICAL LIMITED
Manufacturer StreetLISMEEHAN O'CALLAGHAN'S MILLS
Manufacturer CityCO. CLARE
Manufacturer CountryEI
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameLIPASE
Generic NameLIPASE REAGENT
Product CodeCDT
Date Received2011-05-12
Model NumberNA
Catalog NumberOSR6230
Lot Number1026
ID NumberNA
Device Expiration Date2012-06-01
OperatorHEALTH PROFESSIONAL
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER BIOMEDICAL LIMITED
Manufacturer AddressLISMEEHAN O'CALLAGHAN'S MILLS CO. CLARE, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.