COULTER LH 750 HEMATOLOGY ANALYZER 6605632

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-05-13 for COULTER LH 750 HEMATOLOGY ANALYZER 6605632 manufactured by Beckman Coulter, Inc..

Event Text Entries

[1982258] A customer contacted beckman coulter inc. (bci) stating that the coulter lh 750 analyzer mis-read the last digit of the sample id for a single patient sample in the automatic mode. Sample id (b)(4) was incorrectly read as (b)(4). The mis-read sample id was identified when the instrument generated a "no match" message. No erroneous results were reported out of the laboratory. The sample was rerun and it was read correctly. No death, serious injury, or change to patient treatment is associated with this event.
Patient Sequence No: 1, Text Type: D, B5


[9009728] A field service engineer (fse) was dispatched and reviewed the data with the customer. The fse replaced the laser bar code decoder card and the scanner ribbon cable. The fse ran a full cassette of specimens including the specimen that misread, all read without error. The root cause is unknown; however, the instrument's checksum was disabled. Per labeling, beckman coulter strongly recommends the use of bar-code checksums to provide automatic checks for read accuracy.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2011-00409
MDR Report Key2089072
Report Source06
Date Received2011-05-13
Date of Report2011-04-13
Date of Event2011-04-13
Date Mfgr Received2011-04-13
Device Manufacturer Date2003-02-01
Date Added to Maude2012-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameCOULTER LH 750 HEMATOLOGY ANALYZER
Generic NameAUTOMATED DIFFERENTIAL CELL COUNTER
Product CodeLOQ
Date Received2011-05-13
Model NumberLH 750
Catalog Number6605632
Lot NumberN/A
ID NumberSV 2D1
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address11800 SW 147TH AVENUE MIAMI FL 33196 US 33196


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.