CAPTURE-CMV INDICATOR RED CELLS 0066238

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-05-19 for CAPTURE-CMV INDICATOR RED CELLS 0066238 manufactured by Immucor, Inc..

Event Text Entries

[2005444] Customer reported unexpected negative reactions with capture-cmv on the galileo. When re-testing samples, they found that 6 previously negative results resulted positive when tested with a new lot number of capture-cmv indicator red cells.
Patient Sequence No: 1, Text Type: D, B5


[9121166] A market withdrawal regarding capture-cmv indicator red cells lot 228143, exp 2011-04-29, was released to customers on 11apr2011. The market withdrawal was issued to cover the weaker reactivity that the lot was demonstrating. No confirmatory testing was performed to determine whether results initial results or repeat testing results are correct. No samples were returned to manufacturer for serological testing. Immucor has made several attempts to retrieve images and samples; however, customer is unable to provide the requested file images or additional sample for investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1034569-2011-00057
MDR Report Key2096609
Report Source01,05,06
Date Received2011-05-19
Date of Report2011-05-19
Date of Event2011-04-19
Date Mfgr Received2011-04-19
Device Manufacturer Date2009-08-04
Date Added to Maude2012-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR SCOTT WEBBER
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 30091562
Manufacturer CountryUS
Manufacturer Postal Code30091 5625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-CMV INDICATOR RED CELLS
Generic NameANTI-IGG COATED INDICATOR RED CELLS FOR USE IN SOLID PHASE ASSAYS
Product CodeLJO
Date Received2011-05-19
Catalog Number0066238
Lot Number228143
Device Expiration Date2011-04-29
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 30091562 US 30091 5625


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.