[2006011] A customer contacted beckman coulter stating that the coulter lh 750 slidestainer unit was running when it alerted the user for fluid in the tubing tray. There was some clear liquid (di water) observed in the bottom of the tubing tray. The user was wearing personal protective equipment (ppe) at the time of the event and was not exposed to any biohazard (only di water). There was no report of exposure to mucous membranes or open wounds. No injury was reported and medical attention was not sought.
Patient Sequence No: 1, Text Type: D, B5


[9077874] Bci hotline performed troubleshooting over the phone and operator found a loose line. The customer reconnected the line and cleaned the tubing tray. However, when the bath drain was recycled, the line popped off again dripping di water on to tray. A field service engineer (fse) was dispatched on-site. Fse indicated the waste pump was unplugged and the replaced the plugged waste fitting. Fse verified repair per established procedures and results met published performance specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2011-00435
MDR Report Key2098073
Report Source06
Date Received2011-05-20
Date of Report2011-04-22
Date of Event2011-04-22
Date Mfgr Received2011-04-22
Device Manufacturer Date2008-04-01
Date Added to Maude2012-03-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street11800 SW 147TH AVENUE,
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOULTER? LH 750 SLIDEMAKER
Generic NameSLIDEMAKER
Product CodeKQC
Date Received2011-05-20
Model NumberNA
Catalog Number6605645
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address11800 SW 147TH AVENUE, MIAMI FL 33196 US 33196


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-20