CORTISOL 11875116160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-05-20 for CORTISOL 11875116160 manufactured by Roche Diagnostics.

Event Text Entries

[2000920] The user received questionable cortisol results for an unknown number of patient samples from the cobas e601 analyzer serial number (b)(4). Of the provided data, the results for one patient sample were discrepant and reported outside the laboratory. The initial result was >63. 44 ug/dl and the repeat result on (b)(6) 2011 was 97. 89 ug/dl. No adverse events were alleged regarding the issue. Upon evaluation of the analyzer, the field service representative determined the detection unit was damaged due to fluid. He replaced the detection unit and verified the analyzer operation by running performance testing with results within specifications.
Patient Sequence No: 1, Text Type: D, B5


[9072705] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[9249515] The investigation determined the cause to be the leaking detection unit which was discovered by the field service representative. He found liquid had leaked on a pcb board and changed the board. The issue was solved by the replacement of the whole detection unit. No patients were affected in any way.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2011-02623
MDR Report Key2098145
Report Source05,06
Date Received2011-05-20
Date of Report2011-07-22
Date of Event2011-04-24
Date Mfgr Received2011-04-25
Date Added to Maude2011-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA RAE ANN FARROW
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175213194
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCORTISOL
Generic NameENZYME IMMUNOASSAY, CORTISOL, SALIVARY
Product CodeNHG
Date Received2011-05-20
Model NumberNA
Catalog Number11875116160
Lot Number16130602
ID NumberNA
Device Expiration Date2012-05-31
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-20