CHOLESTEROL GEN.2 03039773190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-05-23 for CHOLESTEROL GEN.2 03039773190 manufactured by Roche Diagnostics.

Event Text Entries

[15724209] The cause of this event was determined to be customer handling of sample tubes. The customer was contacted reagarding proper handling of sample tues, not placing taller sample tubes into racks dedicated to be used with shorter tubes, and pouring short samples into sample cups for testing.
Patient Sequence No: 1, Text Type: N, H10


[17557836] The user stated they were receiving questionable cholesterol results on their cobas c501 (serial number (b)(4)). The user was able to provide data for ten patients. Of that data, there was one discrepant cholesterol result reported outside the laboratory. The repeat was performed on (b)(6) 2011 and sent out on a corrected report. The patient had an initial result of 7 mg/dl the repeat result was 202 mg/dl. There were no adverse affect to the patient as a result of this event. The user declined a service visit. The user believed the problem was caused by a faulty probe that may have been partially occluded by gel. He changed the probe and there have been no further issues.
Patient Sequence No: 1, Text Type: D, B5


[17653452] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2011-02722
MDR Report Key2098720
Report Source05,06
Date Received2011-05-23
Date of Report2011-06-17
Date of Event2011-04-20
Date Mfgr Received2011-04-29
Date Added to Maude2011-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA RAE ANN FARROW
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175213194
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCHOLESTEROL GEN.2
Generic NameENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL
Product CodeCHH
Date Received2011-05-23
Model NumberNA
Catalog Number03039773190
Lot Number63506801
ID NumberNA
Device Expiration Date2011-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-23

© 2026 FDA.report