ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN II 806516100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-05-17 for ANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN II 806516100 manufactured by Bio-rad Medical Diagnostics Gmbh.

Event Text Entries

[2148183] The customer complained a false negative reaction of a patient sample containing an anti-jk(a) with anti-human globulin anti-igg solidscreen ii. We received the sample but not the complained lot of anti-human globulin anti-igg solidscreen ii. Sample was tested with our retention sample of anti-human globulin anti-igg solidscreen ii and screening cells on tango in the quality control laboratory and showed very weak positive reactions with one jk(a) positive screening cell. Furthermore, the sample was tested in the gel method and reacted negative. Additionally, the sample was tested in the tube technique with the enhancement medium polyethylene glycol (peg) and showed weak positive reactions. Evaluation of all tests confirm the weakness of the missed antibody. The correct function of the affected anti-human globulin anti-igg solidscreen ii lot was confirmed by testing different weak antibodies. All positive and negative reactions were correct.
Patient Sequence No: 1, Text Type: D, B5


[9021073] The stn# 125098.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610824-2011-00065
MDR Report Key2104808
Report Source05
Date Received2011-05-17
Date of Report2011-05-17
Date of Event2011-04-16
Date Mfgr Received2011-04-18
Date Added to Maude2012-04-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDR. MARC GORZELLIK
Manufacturer StreetINDUSTRIESTRASSE 1
Manufacturer CityDREIEICH 63303
Manufacturer CountryGM
Manufacturer Postal63303
Manufacturer Phone103801528
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANTI-HUMAN GLOBULIN ANTI-IGG SOLIDSCREEN II
Generic NameAHG ANTI-IGG SOLIDSCREEN
Product CodeDEW
Date Received2011-05-17
Catalog Number806516100
Lot Number7035090-01
Device Expiration Date2012-02-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIO-RAD MEDICAL DIAGNOSTICS GMBH
Manufacturer AddressINDUSTRIESTRASSE 1 DREIEICH D-63303 GM D-63303


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-17