RAPID RESPONSE HUT TABLE 1017

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06,07 report with the FDA on 2011-05-16 for RAPID RESPONSE HUT TABLE 1017 manufactured by Medical Positioning, Inc..

Event Text Entries

[2004750] During a (b)(4) study the pt's head was elevated in a steep reverse trendelenburg position when the pt's heart rate fell to 30 bpm. The staff attempted to move the table to a level position using the electric controls and the table was unresponsive. The staff manually removed the pt from the table and placed the pt flat on the floor. At that time the heart rate recovered without any cpr or other med intervention.
Patient Sequence No: 1, Text Type: D, B5


[9022747] This device is scheduled to be evaluated by mpi once it is returned from the customer. A f/u report will be filed once eval is completed detailing the methods, results, and conclusions of the eval. This device was subject to a field correction action for a similar problem ((b)(4)). Mpi has opened a capa ((b)(4)) on the effectiveness of the field action where the investigation shows that the problem with this table is site specific and not systemic to the other tables involved in the field action. This will be tested and confirmed once the company receives the device from the facility.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932056-2011-00004
MDR Report Key2108204
Report Source04,05,06,07
Date Received2011-05-16
Date of Report2011-05-13
Date of Event2011-05-06
Date Mfgr Received2011-05-06
Device Manufacturer Date2007-07-31
Date Added to Maude2012-06-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOE HURTIG, MGR
Manufacturer Street1717 WASHINGTON
Manufacturer CityKANSAS CITY MO 64108
Manufacturer CountryUS
Manufacturer Postal64108
Manufacturer Phone8165952356
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction NumberZ-1973/1975-2011
Event Type3
Type of Report3

Device Details

Brand NameRAPID RESPONSE HUT TABLE
Generic NameHEAD-UP TILT TABLE
Product CodeLGX
Date Received2011-05-16
Model Number1017
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL POSITIONING, INC.
Manufacturer AddressKANSAS CITY MO 64108 US 64108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-05-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.