ECHO TABLE 1283

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06,07 report with the FDA on 2011-05-16 for ECHO TABLE 1283 manufactured by Medical Positioning, Inc..

Event Text Entries

[2147314] The table was in use and neither a pt nor user-facility staff member were injured during the event, mpi feels that a recurrence of the event with a pt on the product could lead to an adverse event. A critical weld on the product was broken when weight was applied to the table by the placement of inanimate objects on the table. The weld separation caused the head end of the table to fall to the floor. If this had occurred while in use with a pt, the event could have lead to contusions and bruising from such a fall.
Patient Sequence No: 1, Text Type: D, B5


[9026339] Upon eval it was noted that the cause of the problem was a "cold weld" which didn't penetrate properly and was able to separate when weight was applied to the product. Mpi opened (b)(4) to investigate the issue to determine the root cause for the problem, and to determine if a field correction is necessary for the issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932056-2011-00005
MDR Report Key2108205
Report Source04,05,06,07
Date Received2011-05-16
Date of Report2011-05-13
Date of Event2011-02-25
Date Mfgr Received2011-05-06
Device Manufacturer Date2003-10-31
Date Added to Maude2012-06-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOE HURTIG, QUALITY MGR
Manufacturer Street1717 WASHINGTON
Manufacturer CityKANSAS CITY MO 64108
Manufacturer CountryUS
Manufacturer Postal64108
Manufacturer Phone8165952356
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameECHO TABLE
Generic NameECHOCARDIOGRAPHY PROCEDURE TABLE
Product CodeIZZ
Date Received2011-05-16
Model Number1283
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL POSITIONING, INC.
Manufacturer AddressKANSAS CITY MO 64108 US 64108


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.