MICRO-MINIATURE LIGHT WIRE CUTTER ODG-1003

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-04-29 for MICRO-MINIATURE LIGHT WIRE CUTTER ODG-1003 manufactured by Gac International.

Event Text Entries

[18615090] In this event, it was reported that the tip of a micro-miniature light wire cutter separated, resulting in the patient receiving a small cut on the lip. No medical/surgical intervention was required to treat the cut.
Patient Sequence No: 1, Text Type: D, B5


[18750517] While no serious injury resulted in this event, there has been a previous report received in the past two years involving a similar device where this malfunction resulted in a serious injury. Therefore, this event meets the criteria for reportability per 21 cfr part 803. The device was returned and evaluated for dimensional measurements and cutting performance and found to be in specification. Also, a dhr review was conducted; no evidence of manufacturing issues or process deviations was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2086211-2011-00003
MDR Report Key2112790
Report Source05
Date Received2011-04-29
Date of Report2011-04-01
Date Mfgr Received2011-04-01
Date Added to Maude2011-10-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactHELEN LEWIS
Manufacturer Street221 W. PHILA. ST., STE. 60 SUSQUEHANNA COMMERCE CENTER W.
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178457511
Manufacturer G1ORTHODENTAL INTL., INC.
Manufacturer Street2306 M.L. KING, SUITE 1
Manufacturer CityCALEXICO CA 92231
Manufacturer CountryUS
Manufacturer Postal Code92231
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMICRO-MINIATURE LIGHT WIRE CUTTER
Product CodeEJB
Date Received2011-04-29
Returned To Mfg2011-04-26
Catalog NumberODG-1003
Lot Number10-33
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGAC INTERNATIONAL
Manufacturer AddressBOHEMIA NY US


Patients

Patient NumberTreatmentOutcomeDate
10 2011-04-29

© 2026 FDA.report