GERZOG MALLET 7-1/2 LEAD 19-790

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-05-23 for GERZOG MALLET 7-1/2 LEAD 19-790 manufactured by Integra, York - Imiltex.

Event Text Entries

[2021118] Customer reported "mallet heads are pitting severely and shavings are coming off during use. This poses a risk to our pts". Additional info received on (b)(6) 2011 from materials manager who reported the mallets are used for hardware implantation and removal as well as striking other instruments such as osteotomes. There were no known injuries. However, the surgeon was concerned that with continued use, bits of metal can only become dislodged from the mallet head and get into the pt's wound.
Patient Sequence No: 1, Text Type: D, B5


[8984803] The device involved in the reported incident is not available for eval. An investigation has been completed based on the reported info. Part number 19-790 gerzog mallet 7-1/2" (head 8oz) is a lead filled mallet with a stainless steel jacket. The striking surface of the mallet is filled with lead which is a soft metal and will be impacted by normal use. Depending on how the striking surface is making contact not only will the lead be effected but lead shavings could possibly be dislodged. Based on the reported info and the investigation results provided, the root cause is unconfirmed due to no return of device. Integra considers this complaint closed. Future incidents of this nature will be documented for recurrence and trending purposes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523190-2011-00036
MDR Report Key2121899
Report Source05,06
Date Received2011-05-23
Date of Report2011-05-23
Date Mfgr Received2011-02-24
Date Added to Maude2011-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAREN FERNANDEZ
Manufacturer Street315 ENTERPRISE DR
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362341
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGERZOG MALLET 7-1/2 LEAD
Generic NameM10 - ORTHOPEDIC
Product CodeHXL
Date Received2011-05-23
Catalog Number19-790
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA, YORK - IMILTEX
Manufacturer AddressYORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2011-05-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.