IMMAGE IMMUNOCHEMISTRY SYSTEM 470000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-06-17 for IMMAGE IMMUNOCHEMISTRY SYSTEM 470000 manufactured by Beckman Coulter Inc..

Event Text Entries

[2052473] A customer contacted beckman coulter inc. (bci) in regards to erroneous high prealbumin (pab) results generated by the immage immunochemistry system. The results were reported out of the laboratory. The samples were repeated and lower results were obtained. Patient treatment was not impacted by this event.
Patient Sequence No: 1, Text Type: D, B5


[9018139] Sample information was not provided. Calibration results from (b)(4) 2011 were acceptable. Qc results on (b)(4) 2011 were out of range high for igg and pab for level 1. Repeat testing brought the results back within range. No other chemistry issues or system errors were noted. On (b)(4) 2011 a bci field service engineer (fse) inspected the instrument and noted qc is working properly. No signs of hardware issues were noted. Advised the customer to run pab qc at the start of the day and at the end f the patient run. No additional issues were reported or noted. Issue was resolved by recalibrating.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-02264
MDR Report Key2132116
Report Source06
Date Received2011-06-17
Date of Report2011-05-17
Date of Event2011-05-14
Date Mfgr Received2011-05-17
Device Manufacturer Date2011-02-27
Date Added to Maude2012-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER/IMMUNOCHEMISTRY SYSTEMS
Product CodeJQX
Date Received2011-06-17
Model NumberIMMAGE
Catalog Number470000
Lot NumberN/A
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.