SUPPRELIN 50MG KIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-06-13 for SUPPRELIN 50MG KIT manufactured by Endo Pharmaceuticals.

Event Text Entries

[2057882] The supprelin implant was placed in the delivery device and tested prior to use. The physician tunnelled the device in the subcutaneous tissue. The physician was unable to withdraw the plunger, a step that is necessary to leave the implant in the tissue. The physician withdrew the delivery device from the pt's arm. The physician still could not withdraw the plunger. The physician grasped the implant and removed it from the delivery device. The end of the implant had been sheared off by the delivery device. The implant was not useable. The implant was leaking fluid through the end that had been trapped in the delivery device. The physician contacted the drug representative regarding this event. The physician had previously complained to the drug representative regarding the poor manufacturing quality of the delivery device. Diagnosis or reason for use: precocious puberty.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5020974
MDR Report Key2132621
Date Received2011-06-13
Date of Report2011-06-13
Date of Event2011-06-02
Date Added to Maude2011-06-21
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameSUPPRELIN 50MG KIT
Generic NameIMPLANT DELIVERY DEVICE
Product CodeFGY
Date Received2011-06-13
Lot Number6076914
ID Number67979-0002-01
Device Expiration Date2012-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerENDO PHARMACEUTICALS


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-06-13

© 2026 FDA.report