COULTER LH 750 ANALYZER 6605632

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-06-20 for COULTER LH 750 ANALYZER 6605632 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2031364] A customer contacted beckman coulter inc. (bci) reporting a blood fluid leak (5ml) in the drip tray under the manual sampling probe of the coulter lh 750 analyzer. The user was wearing personal protective equipment (ppe) consisting of lab coat, gloves and goggles at the time of the incident. No injuries occurred and medical attention was not sought. There was no report of exposure to mucous membranes or open wounds. There was no death, injury or change to patient treatment attributed or associated to this complaint.
Patient Sequence No: 1, Text Type: D, B5


[9014472] A field service engineer (fse) was dispatched on (b)(4) 2011 and indicated a leak at the manual mode probe. To resolve this issue the manual mode probe and the fluid catch box (drip tray) were adjusted by the fse. Repairs per established procedures were verified and results meet published performance specifications. The root cause for the leak was associated to the manual mode probe. Per labeling, beckman coulter, inc. Urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eyewear, gloves, and suitable laboratory attire when operating or maintaining this or any other automated laboratory analyzer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2011-00712
MDR Report Key2133175
Report Source06
Date Received2011-06-20
Date of Report2011-05-20
Date of Event2011-05-20
Date Mfgr Received2011-05-20
Device Manufacturer Date2002-11-01
Date Added to Maude2012-04-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameCOULTER LH 750 ANALYZER
Generic NameAUTOMATED DIFFERENTIAL CELL COUNTER
Product CodeLOQ
Date Received2011-06-20
Model NumberLH 750
Catalog Number6605632
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address11800 SW 147TH AVENUE MIAMI FL 33196 US 33196


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.