COD CURETTE, BRUNS 0, CUP 3.7MM 32-3473

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-06-13 for COD CURETTE, BRUNS 0, CUP 3.7MM 32-3473 manufactured by Codman & Shurtleff, Inc..

Event Text Entries

[21639935] Medwatch report explained that the #0 curette broke off in pt. Broken piece was removed. Incision has to be enlarged because it broke.
Patient Sequence No: 1, Text Type: D, B5


[21870858] Upon completion of the investigation it was noted that the shaft is broken 3/4 of an inch from the distal end. This type of breakage typically occurs when there has been a small crack in the material. It was also noted that there was evidence of corrosion on at least part of the fractured surface. The tip is dull, which suggests excessive use over a prolonged period of time. This instrument is more than 5 years old. It is likely that atypical force was applied to the instrument to compensate for the dull cutting surfaces and resulted in breakage. This however could not be confirmed. It is not possible to determine how the original crack may have formed. This is the first complaint of this type for this product code, therefore, it is considered to be an isolated incident. Based on the results of this investigation no further action is required. Trends will be monitored for this and similar complaints. At the present time this complaint is closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226348-2011-00215
MDR Report Key2135042
Report Source05,06
Date Received2011-06-13
Date of Event2011-05-04
Date Reported to Mfgr2011-06-08
Date Mfgr Received2011-05-17
Date Added to Maude2011-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMATTHEW KING
Manufacturer Street325 PARAMOUNT DR.
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088283106
Manufacturer G1CODMAN & SHURTLEFF, INC.
Manufacturer Street325 PARAMOUNT DR.
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal Code02767
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOD CURETTE, BRUNS 0, CUP 3.7MM
Generic NameINSTRUMENTS
Product CodeFZS
Date Received2011-06-13
Returned To Mfg2011-05-20
Model NumberNA
Catalog Number32-3473
Lot Number056
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCODMAN & SHURTLEFF, INC.
Manufacturer Address325 PARAMOUNT DR. RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-06-13