IMMAGE 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-06-23 for IMMAGE 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[2117361] A customer contacted beckman coulter inc. (bci) in regards to an erroneous low microalbumin (ma) two patients generated by the image 800 immunochemistry. The erroneous results were reported out of the laboratory. The initial patient samples were repeated and higher results were obtained. Amended reports were initiated. The customer has not received any reports of affects to patients due to the incorrect results.
Patient Sequence No: 1, Text Type: D, B5


[9166955] The samples were urine. The customer did not note any sample or system issue(s). On (b)(4) 2011, a bci field service engineer (fse) found air in sample probe tube and wash station filled. Fse flushed both proves thru valves. Steam was emitting from the reagent probe, which indicates a blockage. Fse aligned and replaced the cuvette wash head teflon which was coated with dried buffer. Replaced filter on cuvette wash drain line. Igg precision run at 2. 04% cv. Instrument is operating acceptably now.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-02219
MDR Report Key2137636
Report Source06
Date Received2011-06-23
Date of Report2011-03-30
Date of Event2011-03-26
Date Mfgr Received2011-03-30
Device Manufacturer Date2007-12-24
Date Added to Maude2012-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Remedial ActionPM
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER/IMMUNOCHEMISTRY SYSTEMS
Product CodeJQX
Date Received2011-06-23
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberN/A
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.