KARL STORZ 11573A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2011-06-18 for KARL STORZ 11573A manufactured by Karl Storz Gmbh & Co. Kg.

Event Text Entries

[16730360] Allegedly, during a sialoendoscopy procedure, doctor stated scope visualization was inadequate, so he proceeded with an open procedure and removed a salivary gland that he felt he had to remove anyway. Patient condition post op was good.
Patient Sequence No: 1, Text Type: D, B5


[17043369] The semi-rigid scope has broken light fibers resulting in inadequate visualization capability. The eyepiece is leaking, the cable is leaking at where it is attached to the scope. The image bundle as well as the light bundle were compromised by the leak.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610617-2011-00025
MDR Report Key2139625
Report Source05
Date Received2011-06-18
Date of Report2011-06-16
Date of Event2011-05-18
Date Facility Aware2011-05-18
Report Date2011-06-16
Date Reported to FDA2011-06-16
Date Reported to Mfgr2011-06-16
Device Manufacturer Date2010-03-01
Date Added to Maude2012-06-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMITTELSTRASSE 8 POSTFACH 230
Manufacturer CityTUTTLINGEN 78503
Manufacturer CountryGM
Manufacturer Postal78503
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKARL STORZ
Generic NameSIALENDOSCOPY TELESCOPE
Product CodeFBP
Date Received2011-06-18
Model Number11573A
Catalog Number11573A
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age15 MO
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ GMBH & CO. KG
Manufacturer AddressTUTTLINGEN GM


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-06-18

© 2026 FDA.report