[2114218] Olympus was informed that during a therapeutic bilateral arthrolectomy procedure, there was a complete loss of image. The procedure was completed with a different but similar device. There was no pt harm reported.
Patient Sequence No: 1, Text Type: D, B5


[9167061] Olympus followed-up with the user facility to gather additional info regarding this report. The user facility reported that the subject device was functioning properly during the first half of the procedure, however, when the subject device reached the vasculature and the surgeon was cauterizing vessels, the image became intermittent, and was subsequently lost. The device referenced in this report was returned to olympus for eval. The eval confirmed the user's report. The image was observed to flicker intermittently when the outer tube was tapped. A severe dent was noted on the outer tube which appears to have caused or contributed to the reported event. The cause of the dent was attributed to mishandling. The device has now been refurnished. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2011-00015
MDR Report Key2140392
Report Source06,07
Date Received2011-06-17
Date of Report2011-05-20
Date of Event2011-05-20
Date Mfgr Received2011-05-20
Date Added to Maude2011-08-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2011-06-17
Returned To Mfg2011-05-25
Model NumberA50021A
Catalog NumberUNK
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKEUHNSTRASSE 61 HAMBURG 22045 GM 22045