TOTAL PROTEIN 12000903216

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-06-28 for TOTAL PROTEIN 12000903216 manufactured by Roche Diagnostics.

Event Text Entries

[17410613] The user received a questionable total protein result for one patient sample from the analytical p module analyzer serial number (b)(4). The initial result from an aliquot of the sample was 11. 8 g/dl and was reported outside the laboratory. The original sample was repeated and the result was 6. 6 g/dl. The aliquot was then repeated and the result was 6. 6 g/dl. The reported result was corrected on (b)(6) 2011. The user did not have access to information concerning if the patient was adversely affected. The total protein r2 reagent lot number was 63554801. Upon evaluation of the analyzer, the field service representative could not determine a cause. He checked multiple areas of the analyzer and all were acceptable. To verify the analyzer operation, he ran precision testing on the total protein assay.
Patient Sequence No: 1, Text Type: D, B5


[17656419] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2011-03499
MDR Report Key2143257
Report Source05,06
Date Received2011-06-28
Date of Report2011-06-28
Date of Event2011-06-14
Date Mfgr Received2011-06-15
Date Added to Maude2011-06-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PROTEIN
Generic NameBIURET (COLORIMETRIC), TOTAL PROTEIN
Product CodeCEK
Date Received2011-06-28
Model NumberNA
Catalog Number12000903216
Lot Number63476401
ID NumberNA
Device Expiration Date2012-02-29
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-28