NUSTAR HYDRAULIC CHAIR DC 1700 DC1700

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2011-06-02 for NUSTAR HYDRAULIC CHAIR DC 1700 DC1700 manufactured by Dental Equipment, Llc (dba Newberg Mfg).

Event Text Entries

[2116189] A dr office allegedly claimed they saw fire coming out of the side of the dental chair when the chair is activated.
Patient Sequence No: 1, Text Type: D, B5


[9168445] During the eval of the dental chair by an authorized svc tech it was observed that someone had installed an after market monitor on the chair and had routed the monitor power cable through the inside of the dental chair after the chair had been installed. The spring mechanism inside the seat area of the chair had rubbed through the outer insulation of the power cord which resulted in sparks being emitted from the chair every time the spring assembly rubbed against the monitor power cable. There was no evidence of fire during the eval of the dental chair as there as no smoke damage, soot, or fire damage inside or outside the dental chair. Upon further investigation it was discovered that an unauthorized it svc tech had installed the after market monitor.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004517290-2011-00006
MDR Report Key2145311
Report Source08
Date Received2011-06-02
Date of Report2011-05-03
Date of Event2011-05-03
Date Mfgr Received2011-05-03
Device Manufacturer Date2009-10-01
Date Added to Maude2011-07-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MGR.
Manufacturer Street11727 FRUEHAUF DR
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1DENTAL EQUIPMENT, LLC DBA NEWBERG MFG
Manufacturer Street705 S. SPRINGBROOK RD. BLDG C
Manufacturer CityNEWBERG OR 97132
Manufacturer CountryUS
Manufacturer Postal Code97132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNUSTAR HYDRAULIC CHAIR DC 1700
Generic NameDENTAL CHAIR
Product CodeKLC
Date Received2011-06-02
Model NumberDC1700
Catalog NumberDC1700
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDENTAL EQUIPMENT, LLC (DBA NEWBERG MFG)
Manufacturer Address705 S. SPRINGBROOK RD. BLDG C NEWBERG OR 97132 US 97132


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-02

© 2026 FDA.report