MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-05-26 for CARDIOCAP 5 manufactured by Ge Healthcare Finland Oy.
[15629603]
It was reported that the cardiocap 5 pt monitor emitted a sizzling sound and stopped functioning (not powering up). It was found that one of the capacitors in the ac/dc power supply was leaking. There was no report of pt involvement or harm to the user. Ge healthcare's investigation is ongoing. A follow up report will be submitted once the investigation has been completed.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 9610105-2011-00012 |
| MDR Report Key | 2145598 |
| Report Source | 05,06 |
| Date Received | 2011-05-26 |
| Date of Report | 2011-05-26 |
| Date of Event | 2011-04-27 |
| Date Mfgr Received | 2011-04-28 |
| Device Manufacturer Date | 2006-09-01 |
| Date Added to Maude | 2012-03-23 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Reporter Occupation | BIOMEDICAL ENGINEER |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 0 |
| Event Location | 0 |
| Manufacturer Contact | JOY SONSALLA |
| Manufacturer Street | 3000 N. GRANDVIEW BLVD. W450 |
| Manufacturer City | WAUKESHA WI 53188 |
| Manufacturer Country | US |
| Manufacturer Postal | 53188 |
| Manufacturer Phone | 2625482661 |
| Manufacturer G1 | GE HEALTHCARE FINLAND OY |
| Manufacturer Street | KUORTANEENKATU 2 |
| Manufacturer City | HELSINKI |
| Manufacturer Country | FI |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | CARDIOCAP 5 |
| Generic Name | MULTI-PARAMETER PATIENT MONITOR |
| Product Code | MLD |
| Date Received | 2011-05-26 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Age | DA |
| Device Eval'ed by Mfgr | N |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | GE HEALTHCARE FINLAND OY |
| Manufacturer Address | HELSINKI FI |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2011-05-26 |