IMMAGE? IMMUNOCHEMISTRY SYSTEM 470000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-07-01 for IMMAGE? IMMUNOCHEMISTRY SYSTEM 470000 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2070095] The customer reported that on (b)(6) 2011 erroneous, high prealbumin (pab) results were generated on an immage immunochemistry system for three patient samples. Subsequent repeat pab results were lower, either being within or below normal reference range. The erroneous results were not reported out of the laboratory and hence there were no deaths, serious injuries or change to patients' treatment associated or attributed to this event. Samples were serum. Pab quality control (qc) results during the timeframe of the event were within established specification range and assay calibration results prior to the event met specification. There were no other chemistry or system errors at the time of the event.
Patient Sequence No: 1, Text Type: D, B5


[9282660] Service was dispatched to the site for this event. The field service engineer (fse) replaced the cuvettes, cuvette wash station, reagent probe motor, reagent arm and performed necessary instrument alignments. Although the fse repaired the instrument prior to returning it into service, a root cause for this event cannot be determined to date.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-02578
MDR Report Key2147677
Report Source05,06
Date Received2011-07-01
Date of Report2011-06-09
Date of Event2011-06-09
Date Mfgr Received2011-06-09
Device Manufacturer Date2001-02-27
Date Added to Maude2012-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-07-01
Model NumberNA
Catalog Number470000
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD. BREA CA 92821 US 92821