IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-07-02 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[17197127] A customer reported to beckman coulter, inc. (bec) a fluid leak on immage 800 immunochemistry system. The customer has a hazard management plan in place. There was no effect to patient or user.
Patient Sequence No: 1, Text Type: D, B5


[17320832] Service visited on (b)(4) 2011, and the field service engineer (fse) found wash head filter occluded with some black material and wash head overflowing cuvettes. The fse replaced filter and performed wash cuvettes, which resulted in no leak or overflow. The fse cleaned up spillage and replaced cuvettes as they were saturated with wash solution externally. There was no further instrument leak complaint filed as of (b)(4) 2011.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-02537
MDR Report Key2148219
Report Source06
Date Received2011-07-02
Date of Report2011-06-02
Date of Event2011-06-02
Date Mfgr Received2011-06-02
Device Manufacturer Date2010-12-02
Date Added to Maude2012-03-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-07-02
Catalog NumberA15445
Lot NumberNA
ID NumberSW 2.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.