LIGHTCYCLER 2.0 03531414001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-07-06 for LIGHTCYCLER 2.0 03531414001 manufactured by Roche Diagnostics.

Event Text Entries

[2101640] The customer stated that upon entering the laboratory she smelled a strong odor resembling an electrical fire. No smoke was observed. The customer identified the source as coming from their lightcycler 2. 0 (serial number (b)(4)). Security investigated the laboratory and determined there was no electrical fire. The customer did not believe there had been a power surge as none of the other equipment plugged into the same wall had been affected. The customer stated no one was injured. Investigation of the event determined the fan to the power supply stopped working and this caused the power supply to overheat, as a result, the electronic elements burnt and the control unit stopped working. No additional complaints of this nature have been received related to the control unit. The issue is detectable as control unit will stop working immediately. The components are ul certified.
Patient Sequence No: 1, Text Type: D, B5


[9196321] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2011-03631
MDR Report Key2151776
Report Source05,06
Date Received2011-07-06
Date of Report2011-07-06
Date of Event2011-03-11
Date Mfgr Received2011-05-03
Date Added to Maude2011-07-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetNONNENWALD 2 NA
Manufacturer CityPENZBERG 82377
Manufacturer Postal Code82377
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameLIGHTCYCLER 2.0
Generic NameCHEMISTRY/MICRO ANALYZER
Product CodeJJF
Date Received2011-07-06
Model NumberNA
Catalog Number03531414001
Lot NumberNA
ID NumberNA
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.