CLINITEK STATUS ANALYZER CT STATUS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-06-29 for CLINITEK STATUS ANALYZER CT STATUS manufactured by Siemens Medical Solutions Diagnostics, Ltd..

Event Text Entries

[2100794] Customer reported a positive urine hcg on a pt. The physician questioned the positive result. The sample was rerun with a negative result. No unnecessary medical procedure was performed. No treatment was withheld as a result of this incident. There was no impact to pt health.
Patient Sequence No: 1, Text Type: D, B5


[9201447] Mfr has contacted customer and requested the reagent lot be returned for investigation. Instrument was also returned and inspected. Instrument failed testing. Optics observed to be out of alignment. Customer is operating with replacement lot of reagent while returned product is investigated by mfr.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2011-00022
MDR Report Key2154507
Report Source07
Date Received2011-06-29
Date of Report2011-06-10
Date of Event2011-06-10
Date Mfgr Received2011-06-10
Device Manufacturer Date2008-01-01
Date Added to Maude2011-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK SMITH
Manufacturer Street2 EDGEWATER DR.
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693158
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MFG. LTD.
Manufacturer StreetNORTHERN RD.
Manufacturer CitySUDBURY, SUFFOLK CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINITEK STATUS ANALYZER
Generic NameCLINITEK STATUS ANALYZER
Product CodeJHJ
Date Received2011-06-29
Model NumberCT STATUS
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS MEDICAL SOLUTIONS DIAGNOSTICS, LTD.
Manufacturer AddressNORTHERN RD. SUDBURY, SUFFOLK CO106DX UK CO10 6DX


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-29