OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2011-06-30 for OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160L manufactured by Olympus Medical System Corporation.

Event Text Entries

[2123353] Olympus was informed that during an unspecified type of colonoscopy procedure, the image quality developed lines and snow to the extent the users could no longer visualize the patient's anatomical structures. The intended procedure was completed with a different olympus endoscope. There was no patient harm reported.
Patient Sequence No: 1, Text Type: D, B5


[9192940] Olympus followed-up with the user facility to gather additional information regarding this report. The user facility reported that the image became intermittent, developed horizontal and vertical lines, and the remaining image was snowy when the subject device reached the cecum. The users reportedly attempted to troubleshoot without success. The device reference in this report was returned to olympus for evaluation. The evaluation confirmed that the image on the scope was found flickering intermittently when the bending section was manipulated. The cause of the phenomenon was isolated to the ccd unit. The unit was serviced and returned to the user facility. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2011-00144
MDR Report Key2160493
Report Source06,07
Date Received2011-06-30
Date of Report2011-06-01
Date of Event2011-06-01
Date Mfgr Received2011-06-01
Date Added to Maude2011-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEM CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS EVIS EXERA COLONOVIDEOSCOPE
Generic NameCOLONOSCOPE
Product CodeFTJ
Date Received2011-06-30
Returned To Mfg2011-06-01
Model NumberCF-Q160L
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEM CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2011-06-30

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.