THERABATH PARAFFIN THERAPY BATH 50 - 60 HZ MODEL TB5 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2011-07-11 for THERABATH PARAFFIN THERAPY BATH 50 - 60 HZ MODEL TB5 * manufactured by Wr Medical Electronics.

Event Text Entries

[2098330] A patient was using the paraffin wax machine and developed a burn on distal tip of middle finger of right hand. The therapist checked the bottom on the inside of the machine and identified that there were several hot spots on the edges along the base (the wax was not too hot). The burn was a very small blister which did not require any treatment. Ten days after the occurrence, a follow up call elicited the information from the patient that all was fine and his finger was healed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2163292
MDR Report Key2163292
Date Received2011-07-11
Date of Report2011-07-11
Date of Event2011-03-02
Report Date2011-07-11
Date Reported to FDA2011-07-11
Date Added to Maude2011-07-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameTHERABATH PARAFFIN THERAPY BATH
Generic NamePARAFFIN WAX BATH
Product CodeIMC
Date Received2011-07-11
Model Number50 - 60 HZ MODEL TB5
Catalog Number*
Lot Number*
ID Number*
OperatorOTHER
Device AvailabilityY
Device Age15 + YR
Device Sequence No1
Device Event Key0
ManufacturerWR MEDICAL ELECTRONICS
Manufacturer Address123 N 2ND ST. STILLWATER MN 55082 US 55082


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-11