RINGLOC 10 DEGREE TRIAL LINER 3/8" SCREW SIZE 24 X 36MM ID N/A 33-117844

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-07-26 for RINGLOC 10 DEGREE TRIAL LINER 3/8" SCREW SIZE 24 X 36MM ID N/A 33-117844 manufactured by Biomet Orthopedics.

Event Text Entries

[18123096] It was reported that patient underwent hip arthroplasty procedure in (b)(6) 2011. During the procedure, after impaction of the acetabular cup, the surgeon inserted the trial liner and after fixating the liner to the cup he noted that the screw portion had separated from the trial liner. The surgeon removed the screw from the bottom of the cup and continued with the procedure. There was no injury to the patient or delay in the procedure as a result.
Patient Sequence No: 1, Text Type: D, B5


[18452158] Review of device history records show that lot released with no recorded anomaly or deviation. Date of event: (b)(6) 2011, exact date unknown. Evaluation of the returned component found evidence to suggest that it failed via a push-out overload of the fastener through the apical hole of the provisional liner. The user facility was notified of the event on (b)(6), 2011. To date, a response has not been received from the user facility. In the event that the user facility submits a medwatch report, biomet will forward a copy to the fda. This report filed (b)(4), 2011.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1825034-2011-00632
MDR Report Key2179320
Report Source07
Date Received2011-07-26
Date of Report2011-06-29
Date Mfgr Received2011-06-29
Device Manufacturer Date2010-06-02
Date Added to Maude2011-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. CRAIG BUCHMAN
Manufacturer Street56 EAST BELL DRIVE P.O. BOX 587
Manufacturer CityWARSAW IN 465810587
Manufacturer CountryUS
Manufacturer Postal465810587
Manufacturer Phone5743711132
Manufacturer G1BIOMET ORTHOPEDICS
Manufacturer Street56 EAST BELL DRIVE P.O. BOX 587
Manufacturer CityWARSAW IN 46581058
Manufacturer CountryUS
Manufacturer Postal Code46581 0587
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRINGLOC 10 DEGREE TRIAL LINER 3/8" SCREW SIZE 24 X 36MM ID
Generic NameDEVICE, PROSTHESIS ALIGNMENT
Product CodeIQO
Date Received2011-07-26
Returned To Mfg2011-06-29
Model NumberN/A
Catalog Number33-117844
Lot Number21443551
ID NumberN/A
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMET ORTHOPEDICS
Manufacturer Address56 EAST BELL DRIVE P.O. BOX 587 WARSAW IN 46581058 US 46581 0587


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.