[2040594] Beckman coulter inc. (bec) (b)(4) reported receiving three (3) damaged ostase calibrator, which leaked. Beckman coulter (b)(4) is not questioning any patient results. No injury or exposure to the fluid was reported.
Patient Sequence No: 1, Text Type: D, B5


[9279371] An investigation of the damaged ostase qc and calibrator vials was conducted at bec (b)(4). The investigation determined the glass vials are not compatible to freezing at -70 degrees celsius. A 100% inspection is conducted when the vial labeling process is performed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2122870-2011-02584
MDR Report Key2184842
Report Source01,05,06
Date Received2011-07-30
Date of Report2011-06-29
Date of Event2011-06-29
Date Mfgr Received2011-06-29
Device Manufacturer Date2011-02-01
Date Added to Maude2012-06-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street1000 LAKE HAZELTINE DRIVE
Manufacturer CityCHASKA MN 55318
Manufacturer CountryUS
Manufacturer Postal Code55318
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction NumberZ-2605-6-2011
Event Type3
Type of Report3

Device Details

Brand NameACCESS OSTASE IMMUNOENZYMETRIC ASSAY
Generic NameCALIBRATOR
Product CodeJIS
Date Received2011-07-30
Model NumberOSTASE CALIBRATOR
Catalog Number37305
Lot Number114260
ID NumberN/A
Device Expiration Date2011-11-30
OperatorMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address1000 LAKE HAZELTINE DRIVE CHASKA MN 55318 US 55318


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-30