THERASEED PALLADIUM 103 ISOTOPE *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1999-04-14 for THERASEED PALLADIUM 103 ISOTOPE * manufactured by Theragenics Corp..

Event Text Entries

[127766] Two shipments were received, each having a different lot number. Shipment a contained the calibration seed for shipment b, and shipment b contained the calibration seed for shipment a. If the problem had not been noticed, pts would have been underdosed or overdosed by 30% depending on which lot number they received. The facility used their own calibration seed. The problem was reported to the mfr, but the mfr just said that the wrong calibration seed got in the wrong shipment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1016130
MDR Report Key219121
Date Received1999-04-14
Date of Report1999-04-13
Date Added to Maude1999-04-19
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationPHYSICIST
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameTHERASEED PALLADIUM 103 ISOTOPE
Generic NameIMPLANT FOR PROSTATE CANCER
Product CodeIWG
Date Received1999-04-14
Model Number*
Catalog Number*
Lot Number9914F
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key212505
ManufacturerTHERAGENICS CORP.
Manufacturer Address5325 OAKBROOK PARKWAY NORCROSS GA 30093 US

Device Sequence Number: 2

Brand NameTHERASEED PALLADIUM 103 ISOTOPE
Generic NameIMPLANT FOR PROSTATE CANCER
Product CodeIWG
Date Received1999-04-14
Model Number*
Catalog Number*
Lot Number9914G
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Age*
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key212581
ManufacturerTHERAGENICS CORP.
Manufacturer Address5325 OAKBROOK PKWY NORCROSS GA 30093 US


Patients

Patient NumberTreatmentOutcomeDate
10 1999-04-14