GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2011-07-27 for GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN manufactured by .

Event Text Entries

[2130557] An ignition involving a gce liv oxygen valve (gce combilite i viprs, 5 micron filter, liv for oxygen) occurred on (b)(6) 2011 at (b)(6) hospital centre in (b)(6). No injury to pts or hosp staff was reported. Based on preliminary info, it appears that the ignition occurred when hosp staff was connecting a hose to the quick connector of the valve. During the transfer of the pt through the emergency entrance, the physician picked up a 5 l liv oxygen cylinder from the stock and came back to the ambulance. Inside the ambulance, the doctor removed the tamper proof evidence (the transparent bag), disconnected the respiratory hose from respiratory vehicle supply and opened the shut off valve (sov). He has taken the hose of the respirator device and has "made closer" the quick connection to the low pressure outlet. When the straight hose connection touched the vipr, a spark occurred. Despite the spark, carried away by the "elan", he has plugged but not locked the quick connection to the low pressure outlet. He heard a "vent noise" followed by a brief explosion with a yellow almost white flame. The ignition caused a small burn to the doctor's pants. The hose was ejected from the valve outlet, the cylinder purged itself through the respiratory hose and fell to the floor of the emergency vehicle. The valve and the hose will be shipped to (b)(4) for investigation. Add'l info is expected. (b)(4) is the mfr for the present device in (b)(4). (b)(4) has a 510(k) for a similar product.
Patient Sequence No: 1, Text Type: D, B5


[9215539] Accidental spontaneous ignition in oxygen valve. The device is under investigation and analysis for root cause.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003900188-2011-00028
MDR Report Key2192813
Report Source01,05
Date Received2011-07-27
Date Mfgr Received2011-06-27
Device Manufacturer Date2009-05-01
Date Added to Maude2011-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer Street575 MOUNTAIN AVE.
Manufacturer CityMURRAY HILL NJ 07974
Manufacturer CountryUS
Manufacturer Postal07974
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN
Generic NameNONE
Product CodeECX
Date Received2011-07-27
Lot Number9/896335267401
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-27