DEROYAL INSUFFLATION TUBING WITH .3 MICRO FIT. 28-0211

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2011-07-13 for DEROYAL INSUFFLATION TUBING WITH .3 MICRO FIT. 28-0211 manufactured by Deroyal Industries, Inc..

Event Text Entries

[2062245] Surgeons repeatedly are complaining that the tubing is extremely slow to insufflate. Complaints of defective product and maybe the filters are bad. This has occurred during multiple cases of laparoscopic surgery. Insufflating the abdomen using this tubing was very slow. This slowed the progress of surgery. At least six tubings were pulled and demonstrated the same problem.
Patient Sequence No: 1, Text Type: D, B5


[9212457] Completed with info provided by the user facility. After analysis of the device components and other possible root causes, it was determined that the insufflation tubing reported may become occluded due to housing material. The plasticizer may migrate from the tubing into the filter housing causing an occlusion. Heat has also been determined to precipitate the occlusion. Through extensive testing and research, a decision was made to change the filter housing material to (b)(4). (b)(4) is a stronger, higher impact and more chemically resistant material. It is also less affected by higher temperatures. We are confident this action has eliminated the tubing issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1033554-2011-00003
MDR Report Key2195159
Report Source05,06,07
Date Received2011-07-13
Date of Report2010-12-03
Date of Event2010-12-03
Date Facility Aware2010-12-03
Report Date2010-12-03
Date Mfgr Received2010-12-05
Date Added to Maude2012-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street200 DEBUSK LANE
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8659387828
Manufacturer G1DEROYAL INDUSTRIES, INC.
Manufacturer Street700 MARTIN LUTHER KING, JR. BLVD
Manufacturer CitySANDORD FL 32771
Manufacturer CountryUS
Manufacturer Postal Code32771
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEROYAL INSUFFLATION TUBING WITH .3 MICRO FIT.
Generic NameTUBING WITH FILTER, INSUFFLATION
Product CodeNKC
Date Received2011-07-13
Catalog Number28-0211
Lot Number24085731
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEROYAL INDUSTRIES, INC.
Manufacturer AddressPOWELL TN 37849 US 37849


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2011-07-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.