STAINLESS STEEL BLADE 0940025000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2011-08-03 for STAINLESS STEEL BLADE 0940025000 manufactured by Stryker Ireland Ltd..

Event Text Entries

[21368641] It was reported that while removing a padded fibreglass cast, the pt received a cut on the lower leg. It was also reported that the size of the cut was one inch long and was not severe. It was further reported that medical intervention was not required.
Patient Sequence No: 1, Text Type: D, B5


[21534801] The cast cutter blade associated with this event was returned to the manufacturer for evaluation. It was visually confirmed that all teeth were intact and there was no signs of damage to the blade. The returned blade was measured where possible and all critical specifications were met. Lot number information has not been provided to permit further investigation. The ifu for the cast cutter blade has a warning which states "cut with an up and down motion as shown. Straight cutting may cut or burn the pt". The investigation results indicate that the user may have contributed to this event, however this cannot be confirmed. The root cause is undetermined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616696-2011-00128
MDR Report Key2196642
Report Source07
Date Received2011-08-03
Date of Report2011-07-07
Date of Event2011-07-06
Date Mfgr Received2011-07-07
Date Added to Maude2011-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactUNA BARRY
Manufacturer StreetCARRIGTWOHILL BUSINESS & TECHNOLOGY PARK
Manufacturer CityCARRIGTWOHILL, CORK
Manufacturer CountryEI
Manufacturer Phone214532963
Manufacturer G1STRYKER IRELAND LTD.
Manufacturer StreetCARRIGTWOHILL BUSINESS & TECHNOLOGY PARK
Manufacturer CityCARRIGTWOHILL, CORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTAINLESS STEEL BLADE
Product CodeLGH
Date Received2011-08-03
Returned To Mfg2011-07-18
Catalog Number0940025000
Lot NumberUNK
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER IRELAND LTD.
Manufacturer AddressCARRIGTWOHILL, CORK EI


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-08-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.