SYNCHRON CX3 DELTA 467501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-08-16 for SYNCHRON CX3 DELTA 467501 manufactured by Beckman Coulter Inc..

Event Text Entries

[16571393] On (b)(4) 2011, field service engineer (fse) visited and found an evidence of some form of overflowing; however the issue is no longer occurring. Per fse, cx3 overflow pan showed evidence of creatinine and wash fluid residue. Fse cleaned, secured all waste lines, tubing, and checked trap, cups pumps, and reagents for leaks. No further leaking observed. Primed the unit x80 with no leaks noted. Unable to reproduce issue despite running for close to 100 cycles. A definitive root cause is undetermined. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[16673594] A customer contacted beckman coulter inc (bec) to report fluid leak underneath the synchron cx3 delta. The customer used personal protective equipment to clean the leak. No injury or exposure was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2050012-2011-04569
MDR Report Key2213580
Report Source06
Date Received2011-08-16
Date of Report2011-07-17
Date of Event2011-07-17
Date Mfgr Received2011-07-17
Device Manufacturer Date2006-05-23
Date Added to Maude2012-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameSYNCHRON CX3 DELTA
Generic NameMICRO CHEMISTRY ANALYZER FOR CLINICAL USE
Product CodeJJF
Date Received2011-08-16
Model NumberCX3 DELTA
Catalog Number467501
Lot NumberN/A
ID NumberSW 4.7
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S KRAEMER BOULEVARD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-08-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.