SYNCHRON? SYSTEMS BENZODIAZEPINES (BNZG) REAGENT A18298

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-08-17 for SYNCHRON? SYSTEMS BENZODIAZEPINES (BNZG) REAGENT A18298 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2202479] The customer reported that erroneously positive benzodiazepine (bnzg) results were generated from a unicel dxc 600i synchron access clinical system for an unknown number of patient samples for a specific synchron systems benzodiazepines (bnzg) reagent lot. Instrument bnzg quality controls results recovered as positive for negative qc control materials during the timeframe of this event. No erroneous bnzg results were reported outside of the laboratory hence there was no death, serious injury or modification to patient treatment associated or attributed to this event. Initial results were repeated at a third party laboratory. The repeat results were different from the initial results, and were considered valid. The customer changed synchron systems benzodiazepines (bnzg) reagent lots, and quality control results responded by returning back into established specification range. Data provided by the customer indicates the generation of an initial positive bnzg result, which upon repeat on another instrument generated a negative repeat result. The samples were urine patient samples.
Patient Sequence No: 1, Text Type: D, B5


[9401722] Event appears to be reagent related. Service was not dispatched to the site for this event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-03776
MDR Report Key2213762
Report Source05,06
Date Received2011-08-17
Date of Report2011-07-08
Date of Event2011-07-07
Date Mfgr Received2011-07-08
Device Manufacturer Date2011-04-18
Date Added to Maude2012-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNCHRON? SYSTEMS BENZODIAZEPINES (BNZG) REAGENT
Generic NameENZYME IMMUNOASSAY, BENZODIAZIPINE
Product CodeJXM
Date Received2011-08-17
Model NumberNA
Catalog NumberA18298
Lot NumberM104042
ID NumberNA
Device Expiration Date2012-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-08-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.