ACUITY H77

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2011-07-14 for ACUITY H77 manufactured by Varian Medical Systems Uk Ltd.

Event Text Entries

[15146660] During patient setup, the acuity stand pro display cover (pizza board cover) fell while the gantry was moving from varian iec 90 to 270 degree. There was no serious injury reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[15173588] Method: though still under investigation, varian has determined that a mdr is appropriate as this possible malfunction, should it recur, could potentially cause serious injury. Additional follow-up to this mdr is expected upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8020711-2011-00006
MDR Report Key2224344
Report Source01,05,06
Date Received2011-07-14
Date of Report2011-06-15
Date of Event2011-06-15
Date Mfgr Received2011-06-15
Device Manufacturer Date2010-09-01
Date Added to Maude2012-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHYSICIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL PIGNATARO, MANAGER
Manufacturer Street911 HANSEN WAY M/S C-255
Manufacturer CityPALO ALTO CA 94304
Manufacturer CountryUS
Manufacturer Postal94304
Manufacturer Phone6504246471
Manufacturer G1VARIAN MEDICAL SYSTEMS UK LTD
Manufacturer StreetGATWICK RD.
Manufacturer CityCRAWLEY, WEST SUSSEX RH109RG
Manufacturer CountryUK
Manufacturer Postal CodeRH10 9RG
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACUITY
Generic NameSYSTEM, SIMULATION, RADIATION THERAPY
Product CodeKPQ
Date Received2011-07-14
Model NumberH77
ID NumberACS 3.2.19
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVARIAN MEDICAL SYSTEMS UK LTD
Manufacturer AddressCRAWLEY, WEST SUSSEX UK


Patients

Patient NumberTreatmentOutcomeDate
10 2011-07-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.