MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2011-08-29 for SKIN PREP WIPES 420400 manufactured by Smith & Nephew Wound Management.
[2240641]
This skin prep complaint was received post smith & nephew's remedial action (voluntary recall) to prevent an unreasonable risk of substantial harm to the public health (ref. Recall #3006760724-04-06-2011-001r). Adverse incident ms. (b)(6) called to inform us she had gone into the hospital not feeling well and was given antibiotics. The client then had heart failure. She then went into renal failure. The client states there was infection under flange at one time with leaking and infected.
Patient Sequence No: 1, Text Type: D, B5
[9257652]
Smith & nephew was contacted regarding the above described incident, which was reported as a field complaint for "infection-local/site" and "general illness". The customer sent in some product samples. We were unable to confirm the complaint based on inspection of the returned samples, hence laboratory testing was performed. Both the returned sample(s) and control samples (from stock) of lot 0j85 were analyzed by an independent test laboratory and met finished product specifications with no evidence of microbial contamination found. Batch records for the lot indicate all specifications were met at the time of release and no inconsistencies were noted. An independent medical review concluded there was no correlation between the reported symptoms and the use of skin prep wipes. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3006760724-2011-00040 |
MDR Report Key | 2225920 |
Report Source | 04 |
Date Received | 2011-08-29 |
Date of Report | 2011-05-18 |
Date of Event | 2011-05-18 |
Date Mfgr Received | 2011-05-18 |
Date Added to Maude | 2011-08-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR TERRY MCMAHON |
Manufacturer Street | 970 LAKE CARILLON DRIVE SUITE 110 |
Manufacturer City | ST. PETERSBURG FL 33716 |
Manufacturer Country | US |
Manufacturer Postal | 33716 |
Manufacturer Phone | 7273993785 |
Manufacturer G1 | SMITH & NEPHEW WOUND MANAGEMENT |
Manufacturer Street | 970 LAKE CARILLON DRIVE SUITE 110 |
Manufacturer City | ST. PETERSBURG FL 33716 |
Manufacturer Country | US |
Manufacturer Postal Code | 33716 |
Single Use | 0 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SKIN PREP WIPES |
Generic Name | BANDAGE, LIQUID, SKIN PROTECTANT |
Product Code | NEC |
Date Received | 2011-08-29 |
Model Number | 420400 |
Lot Number | 0J85 |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SMITH & NEPHEW WOUND MANAGEMENT |
Manufacturer Address | 970 LAKE CARILLON DRIVE SUITE 110 ST. PETERSBURG FL 33176 US 33176 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization; 2. Other | 2011-08-29 |