SKIN PREP WIPES 420400

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2011-08-29 for SKIN PREP WIPES 420400 manufactured by Smith & Nephew Wound Management.

Event Text Entries

[2240642] This skin prep complaint was received post smith & nephew's remedial action (voluntary recall) to prevent an unreasonable risk of substantial harm to the public health (ref. Recall #3006760724-04-06-2011-001r). Adverse incident experienced stomach pains and body chills resulting in a hospital stay of four days.
Patient Sequence No: 1, Text Type: D, B5


[9257653] Smith & nephew was contacted regarding the above described incident, which was reported as a field complaint for general illness. The customer sent in some product samples. We were unable to confirm the complaint based on inspection of the returned samples, hence laboratory testing was performed. Both the returned sample(s) of lot 0f164 and control samples of additional lots (from stock) of skin prep wipes were analyzed by an independent test laboratory and met finished product specifications with no evidence of microbial contamination found. Batch records for the lot indicate all specifications were met at the time of release and no inconsistencies were noted. In addition, and independent medical review was performed which concluded that there is no medical evidence to support any relationship between the use of skin-prep wipes and the patient's symptoms. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006760724-2011-00036
MDR Report Key2225922
Report Source04
Date Received2011-08-29
Date of Report2011-08-25
Date of Event2011-05-04
Date Mfgr Received2011-05-04
Date Added to Maude2011-08-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR TERRY MCMAHON
Manufacturer Street970 LAKE CARILLON DRIVE SUITE 110
Manufacturer CityST. PETERSBURG FL 33716
Manufacturer CountryUS
Manufacturer Postal33716
Manufacturer Phone7273993785
Manufacturer G1SMITH & NEPHEW WOUND MANAGEMENT
Manufacturer Street970 LAKE CARILLON DRIVE SUITE 110
Manufacturer CityST. PETERSBURG FL 33716
Manufacturer CountryUS
Manufacturer Postal Code33716
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSKIN PREP WIPES
Generic NameBANDAGE, LIQUID, SKIN PROTECTANT
Product CodeNEC
Date Received2011-08-29
Model Number420400
Lot Number0F164
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW WOUND MANAGEMENT
Manufacturer Address970 LAKE CARILLON DRIVE SUITE 110 ST. PETERSBURG FL 33176 US 33176


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2011-08-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.