IMMAGE IMMUNOCHEMISTRY SYSTEM 470000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2011-09-01 for IMMAGE IMMUNOCHEMISTRY SYSTEM 470000 manufactured by Beckman Coulter Inc..

Event Text Entries

[2324809] A customer contacted beckman coulter inc (bec) to report a reaction wheel temperature error after a preventive maintenance. The customer noted burning smell when cuvette cover was lifted. The customer shut down the instrument. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[9458761] A bec field service engineer (fse) replaced a defective circulation fan in the reaction wheel chamber, which was the cause of the event. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-04696
MDR Report Key2231995
Report Source06
Date Received2011-09-01
Date of Report2011-02-21
Date of Event2011-02-21
Date Mfgr Received2011-02-21
Device Manufacturer Date2001-06-28
Date Added to Maude2012-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER FOR CLINICAL USE
Product CodeJQX
Date Received2011-09-01
Model NumberIMMAGE
Catalog Number470000
Lot NumberN/A
ID NumberSW 1.6
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S KRAEMER BOULEVARD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-01

© 2024 FDA.report