STRUT PICK, .50MM PT.,90 DEG ANG,MED 1/E 3722062

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06,07,company representati report with the FDA on 2011-09-02 for STRUT PICK, .50MM PT.,90 DEG ANG,MED 1/E 3722062 manufactured by Medtronic Xomed Inc..

Event Text Entries

[22123497] It was reported while undergoing an ear procedure, the tip of an oto pick instrument broke off in a patient's ear. Another magnetized instrument was used to retrieve the metal tip. No injury or sequella reported. Long term prognosis was reported as good. The customer did not know if the product broke in the outer or inner ear. It was also reported that the fragment was retrieved quickly and without difficulty.
Patient Sequence No: 1, Text Type: D, B5


[22259749] Patient weight was not known by reporter. The customer did not have the product lot number. Instrument will not be returned to the manufacturer. An oto pick is used for a variety of ear procedures, most commonly in middle ear procedures. It is long and narrow with a very fine tip. Some picks have a slight angle at the tip for better access to the anatomy of the ear.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2011-00064
MDR Report Key2232139
Report Source00,06,07,COMPANY REPRESENTATI
Date Received2011-09-02
Date of Report2011-08-08
Date of Event2011-04-12
Report Date2011-06-09
Date Reported to FDA2011-06-09
Date Mfgr Received2011-08-08
Date Added to Maude2012-06-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMIKE MOSBY
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9042797584
Manufacturer G1MEDTRONIC XOMED INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRUT PICK, .50MM PT.,90 DEG ANG,MED 1/E
Generic NamePICK, MICROSURGICAL EAR
Product CodeJYT
Date Received2011-09-02
Model Number3722062
Catalog Number3722062
Lot NumberNOT PROVID
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INC.
Manufacturer Address6743 SOUTHPOINT DR. N. JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-02

© 2026 FDA.report