[16123146] Pt had a c5-c7 inion plate for an acdf in (b)(6) 2011. The pt developed loss of her lordotic curvature after the surgery on f/u films: reversal of lordosis at c5-7. Partial collapse of c4/5 also. In the revision surgery the inion plate and screws were removed. The screw heads were found loose which according to surgeon may or may not have occurred during dissecting fibrous tissue off the plate. The plate cracked into two during removal. The pt then was reconstructed with a new c4/5 acdf and revision of the c5/6 fusion and placement of a new cage in the c5/6 space in addition to the cage that was already there. Then a metal plate was used from c4-7 to complete the procedure. Surgeon conclusion is loss of bone structure. The pt had been in motor vehicle accident prior to the operation.
Patient Sequence No: 1, Text Type: D, B5


[16244387] Part 6: eval and conclusion will be provided in a f/u report after completion.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9710629-2011-00009
MDR Report Key2236980
Report Source05,08
Date Received2011-08-31
Date of Report2011-08-31
Date of Event2011-06-22
Date Mfgr Received2011-06-30
Device Manufacturer Date2008-04-01
Date Added to Maude2011-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetLAAKARINKATU 2
Manufacturer CityTAMPERE 33520
Manufacturer CountryFI
Manufacturer Postal33520
Manufacturer Phone108306600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINION S-1 BIODEGR. GRAFT CONTAINMENT PLATE
Generic NameGRAFT CONTAINMENT PLATE
Product CodeOJB
Date Received2011-08-31
Model NumberSPN-5017
Catalog NumberSPN-5017
Lot Number0804052
Device Expiration Date2011-04-01
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINION OY
Manufacturer AddressLAAKARINKATU 2 TAMPERE 33520 FI 33520


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2011-08-31