IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-09-07 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2173341] A customer reported to beckman coulter, inc. (bec) that a fluid leak was observed from back of immage 800 immunochemistry system. Msds was not reviewed, but the facility has an exposure control plan. The operator did not seek medical attention and there was no effect to patient or user.
Patient Sequence No: 1, Text Type: D, B5


[9292067] On (b)(4) 2011, field service engineer (fse) found all cuvettes overflowing with wash solution. The fse replaced drain filter, and cleaned all cuvettes. The fse performed wash cuvettes with no flooding, but cuvettes started to flood again upon qc run. The fse found the waste filter was blocked with gel substance inside filter, and replaced the filter and wash probe. The fse performed alignments with new cuvettes in place. The fse also replaced all other filters for both wash stations and hydro. The fse rebooted the instrument and performed wash cuvettes with no flooding. Sample wash station started to overflow. The fse replaced both wash station waste valves, controller 2 board and controller 1 board due to flooding from cuvette wash probe. The fse washed all cuvettes three times, and primed the system five times with no issues. The fse ran qc for all chemistries 2 times. No issue was noted and all qc results were within the established ranges. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-05045
MDR Report Key2237275
Report Source05,06
Date Received2011-09-07
Date of Report2011-08-08
Date of Event2011-08-08
Date Mfgr Received2011-08-08
Device Manufacturer Date2008-08-19
Date Added to Maude2012-05-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149613634
Manufacturer G1BECKMAN COULTER, INC
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2011-09-07
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberNA
ID NumberSW 1.6
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.