IMMAGE? IMMUNOCHEMISTRY SYSTEM 470000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2011-09-09 for IMMAGE? IMMUNOCHEMISTRY SYSTEM 470000 manufactured by Beckman Coulter Inc..

Event Text Entries

[21426377] A customer contacted beckman coulter inc. (bec) stating that the samples and reagent probe wash stations in the immage immunochemistry system were not draining and were overflowing. No injury or exposure was reported and no erroneous results were generated.
Patient Sequence No: 1, Text Type: D, B5


[21535680] The customer performed troubleshooting with bec cts (customer technical support) and noted that the sample and reagent probe wash stations were not draining and overflowing as a result of a lack of vacuum. Cts generated a service request. A field service engineer (fse) went on-site (b)(4) 2011 and replaced a diaphragm due to the low vacuum issue. Repair was verified per established procedures and results met published performance specifications. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2011-05090
MDR Report Key2240188
Report Source05,06
Date Received2011-09-09
Date of Report2011-08-11
Date of Event2011-08-11
Date Mfgr Received2011-08-11
Device Manufacturer Date2010-06-04
Date Added to Maude2012-05-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. NORA ZEROUNIAN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7168282419
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? IMMUNOCHEMISTRY SYSTEM
Generic NameCLINICAL CHEMISTRY, NEPHELOMETER
Product CodeJQX
Date Received2011-09-09
Model NumberIMMAGE
Catalog Number470000
Lot NumberNA
ID NumberSW VERSION 1.6
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2011-09-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.